Room Code | MRI-47-I |
Net Room Area Briefed | 47 |
Hours of Operation | 8am to 10pm, daily |
Occupancy | 1 patient; up to 4 staff assisting |
Description / Special Requirements | Room for MRI scanning procedures, with adjacent control room and computer room. A staff handwashin basin will be located outside the room in close proximity. Fixtures and fittings to be non magnetic. Body Protected electrical area Acoutsics: acoustic privacy required for ambient noise emitted by scanning equipment Special Design Requirements: - The MRI suite shall be planned to conform to staged access and control zones complying with local authority requirements to prevent unscreened persons accessing the 5-gauass volume around the MRI scanner and to minimise radiofrequency interference from other equipment - Electromagnetic sheilding (radio frequency enclosure) as advised by RF Shielding Consultant - Structural assessment may be required to ensure the floor is designed to acommodate the weight of the equipment - Floor to ceiling cable duct with removable covers may be required - Underfloor cable trays between Control area and MRI table and floor to ceiling duct may be required - Warning lights outside each entry door - Power to be on emergency supply - all power to MRI to pass through MRI isolation transformer - Installation of the MRI including services requirements such as cryogen venting, emergency exhaust, passive pressure relief systems, shall conform to local authority requirements and manufacturer's specifications |
Amendment | 01.02.2011 First Draft 15.03.2017 Description & services updated |
ID | Fabric | Material | Finish | Specification | Selection | Remarks |
5010 | Ceiling | Plasterboard | Paint, washable | Flush set, suspended | at 3000mm height | |
2000 | Skirting | Vinyl | Prefinished | Floor vinyl coved, 150mmH | ||
3000 | Wall finish | Paint | Acrylic, washable | |||
7025 | Observation | Lead glass | Clear | Internal window | to Control Room | |
6025 | Cornice | Aluminium | Powdercoat | shadow line | ||
9110 | Doors | Solid Core | Paint | 1 1/2 leaf | patient entry/ exit; 1400mm clear opening | |
1000 | Floor finish | Vinyl | Standard slip resistant | Seamless, coved |
ID | Description | Category | Group | Qty | Selection / Remarks |
150 | Air flowmeter | Furniture/ Fitting | 3 | 1 | to patient services panel; non-magnetic |
1465 | Bracket: suction bottle | Furniture/ Fitting | 2 | 1 | non-magnetic |
3350 | Clock: battery operated | Furniture/ Fitting | 3 | 1 | non-magnetic |
16250 | Oxygen flowmeter | Furniture/ Fitting | 3 | 2 | to patient services panel; non-magnetic |
21950 | Suction adapter | Furniture/ Fitting | 3 | 2 | to patient services panel; non-magnetic |
ID | Description | Grp | Qty | Ele | Data | CdW | HtW | WmW | Tap | Dns | Gas | Selection / Remarks |
7300 | Camera: CCTV security | 1 | 3 | yes | ___ | ___ | ___ | ___ | ___ | ___ | ___ | to view patient, optional |
66750 | IMG: MRI imaging unit | 3 | 1 | yes | yes | ___ | ___ | ___ | ___ | ___ | ___ |
ID | Description | Service Category | Qty | Selection / Remarks |
1006 | Voice / Data outlet: double | Communications | 1 | 1 outlet for telephone |
1040 | Intercom | Communications | 1 | handsfree to Control Room |
5000 | Airconditioning | HVAC | ||
6001 | General: colour corrected | Lighting | ||
6110 | Downlights: dimmable | Lighting | ||
7000 | Oxygen (O2) | Medical Gases | 2 | patient services panel; non magnetic outlet |
7010 | Nitrous Oxide (N2O) | Medical Gases | 1 | patient services panel; non magnetic outlet |
7020 | Medical Air (MA) | Medical Gases | 1 | patient services panel; non magnetic outlet |
7040 | Suction | Medical Gases | 2 | patient services panel; non magnetic outlet |
7070 | Scavenge | Medical Gases | 1 | patient services panel; non magnetic outlet |
8010 | Staff/ Nurse assist call | Nurse Call | 1 | |
8080 | Emergency call | Nurse Call | 1 | |
9010 | GPO: Emergency power, single | Power | 4 | patient services panel |
9011 | GPO: Emergency power, double | Power | 3 | walls |
9040 | Body protected | Power | ||
9060 | GPO: Special: for equipment | Power | 1 | MRI Imaging equipment, Control and computer equipment modules; refer to Manufacturer's specifications |
9060 | GPO: Special: for equipment | Power | 3 | CCTV cameras, hardwired; optional |
11000 | Wall panel | Services Panels | 1 | patient services panel |
12020 | Faraday cage | Shielding | to walls & doors | |
2465 | Special: 'Magnet on' warning lights | Lighting | 1 | to outside of room entry |